Thermo Scientific APEX User’s Guide Page 105
• Custom speed
This setting is used when your product is flowing through the chute either very
fast or very slowly, or because you are using a non-standard angle (that is, an
angle different from the standard angles of 25°, 35°, and 45°). For help setting
up a custom speed-filter setting, please contact Thermo Fisher Scientific—as
described in the “Contacting Thermo Fisher Scientific” section at the end of the
manual.
Establishing Basic Product Parameters
Before putting your detector into service, you must have it learn how to identify your
uncontaminated product. This is done using the detector’s full-product calibration
function, which sets the following basic parameters for your product.
• Phase-angle setting
• Detect level
Doing a Full-Product Calibration
To do a full-product calibration, follow the instructions below.
1) Make sure product is falling through the chute.
2) Highlight your product name in the Main Menu.
3) Press the Go button and page 1 of the product menu appears.
Product Calibrate and Name