Appendix A - Specifications
Thermo Electron Corporation A-2
Table 2 Environmental and electrical
requirements (meets the IEC 1010-1
international safety standard)
Line voltages
100 - 240 VAC ±10%
50 - 60 Hz
80 VA max
Operating environment
The instrument meets the specifications on the
previous page under the following conditions
after a 30-minute warm-up period.
Ambient temperature: 5° to 35° C
(41° F to 95° F)
Relative humidity - 20% - 80% RH
Storage environment
-20° C to 70° C (-4° F to 158° F) Relative
humidity not to exceed 0.040 pounds
moisture per pound of dry air. Allow
instrument to adjust to room temperature for
24 hours after taking it out of storage.
Temperature should be maintained at ±4° F.
Relative humidity should be maintained to ±5%.
Altitude
From below sea level to 2000 meters
(6562 feet)
For indoor use only
Installation Category II
Pollution Degree 2
BioMate 3
Dimensions 330W x 410D x 235H mm (13" x 16" x 9")
Weight 8.6 Kg (19lb)
Warranty 1 year
1
These specifications are valid only when the required environmental conditions (see below) are met.
2
When working in the range of 800 to 1100nm, allow the instrument to warm up for at least an hour.
3
Measured using NIST 930D filters
4
220nm and 340nm measurements with SPECTRONIC Standards (#333150)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo BioMate 3 series and is the answer not in the manual?

Thermo BioMate 3 series Specifications

General IconGeneral
BrandThermo
ModelBioMate 3 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals