EasyManuals Logo

Thermo EPD Mk2 User Manual

Default Icon
115 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
EPD/HB/40521/000
ISSUE 5
EPD TECHNICAL HANDBOOK Page 51
4.6 EPD Internal Processing & Facilities
This section describes the internal processing performed by the EPD and the
facilities provided over the IR communications link.
4.6.1 Memory Characteristics:
Memory Retention: 10-year data retention.
Dose Profile History: Settable interval from two seconds to 35 hours.
Stores transitions of Hp(10) and Hp(0.07) at a
resolution of 1µSv. Store for up to 579 records
for transitions up to 127 µSv or less.
Dose Store (short term): Hp(10) and Hp(0.07). Can be reset after each
controlled area entry, day, month (etc) by health
physics staff or dose management system,
according to local procedures.
Total Dose Store: Separate total dose store.
Special Total Dose Store: 12 entries, e.g. for automatic recording of end
of month dose etc.
Event Log: 23 entries for time recording of alarms etc., for
incident assessments.
ADS Dose Store: Approved Dosimetry Service (Record of Dose)
dose memory area with password protection.
Peak Dose Rates: Peak dose rates with time of occurrence.
Storage Times: 1 second resolution.
Alarm Flags: Various alarm and fault flags.
Timer: Count-down timer configurable up to 1 hour 39
minutes and 59 seconds (1 second resolution).
Return for Read: Programmable warning.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo EPD Mk2 and is the answer not in the manual?

Thermo EPD Mk2 Specifications

General IconGeneral
BrandThermo
ModelEPD Mk2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals