EPD/HB/40521/000
ISSUE 5
EPD TECHNICAL HANDBOOK Page 72
Services Module
Code Text as displayed by
EasyEPD2
EPD
Action
Recommended
Customer Action
Remarks
030 Service Module stCalData
CRC faulty
Reset unit. Return to
supplier if recurs.
031 Service Module
stRateThres CRC faulty
Reset unit. Return to
supplier if recurs.
032 Service Module
stDoseThres CRC faulty
Reset unit. Return to
supplier if recurs.
033 Service Module
stDoseSummary CRC
faulty
Reset unit. Return to
supplier if recurs.
034 Service Module stAdsData
CRC faulty
Reset unit. Return to
supplier if recurs.
035 Service Module Seconds
Timer Array overflow
Reset Reset unit. Return to
supplier if recurs.
This indicates a software
fault.
036 Service Module Fine Timer
Array overflow
Reset Reset unit. Return to
supplier if recurs.
This indicates a software
fault.
037 Service Module Long Timer
Array overflow
Reset Reset unit. Return to
supplier if recurs.
This indicates a software
fault.
User I/F Module
Code Text as displayed by
EasyEPD2
EPD
Action
Recommended
Customer Action
Remarks
040 User Interface Module Error Reserved for future use
Sounder Module
Code Text as displayed by
EasyEPD2
EPD
Action
Recommended
Customer Action
Remarks
050 Sounder Module Error Reserved for future use