EasyManua.ls Logo

ThermoFisher Scientific HyPerforma DynaDrive - Page 38

ThermoFisher Scientific HyPerforma DynaDrive
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 | Replacing parts
3. Fasten the load cell using a 3/16 in. Allen wrench with the two bolts (Figure 3.53).
Note: Make sure the arrow on the load cell is pointing down.
Figure 3.53. Bolts on top of the load cell.
Two 3/16 in. bolts
4. Using a 1/8 in. Allen wrench, replace the two bolts that connect the load cell to the
base of the leg (Figure 3.54).
Figure 3.54. Replacing the bolts securing the load
cell to a leg.
5. Using a crescent wrench, lower the leg by twisting the lock-out nut counter-
clockwise until weight is on the load cell (Figure 3.55). Note: Lower the nut back
down to remove weight o load cell to repeat with the other legs, if necessary.
Figure 3.55. Twisting load cell lock-out
nut counter-clockwise.
Thermo Scientific 34DynaDrive S.U.B. Service Guide

Table of Contents