Chapter 6 | DynaDrive S.U.B. specifications
Table 6.4. 500 L DynaDrive S.U.B. specifications (continued).
Specification
Support container
Overall width
141 cm (56 in.) with optional E-Box
103 cm (41 in.) without E-Box
Overall length 101 cm (40 in.)
Overall height 311 cm (122 in.)
Dry skid weight (mass)
522 kg (1,151 lb.) with optional E-Box
481 kg (1,060 lb.) without E-Box
Wet skid weight—rated working volume (mass)
1,022 kg (2,251 lb.) with optional E-Box
981 kg (2,160 lb.) without E-Box
General
Electrical power supply requirement (voltage, phase, amp) 120/208–240 VAC, single, 20 A (50/60 Hz)
pH & DO probe—autoclavable and single-use options available
(single-use are integrated into the BPC)
12 mm diameter x 215–235 mm insertion
length x 13.5 PG (pipe) thread
Noise level < 70 dB at 1.5 m
Recommended operating parameters
Operating temperature range 17°C to 40°C ± 0.5°C (104°F ± 0.9°F)
Motor speed 35–120 rpm
Volume range 25– 500 L
Maximum bag pressure 34 mbar (0.5 psi)
Thermo Scientific 67DynaDrive S.U.B. Service Guide