EasyManuals Logo

THORLABS LDC200C series User Manual

THORLABS LDC200C series
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
© 2015 Thorlabs
5 Appendix
25
5.1.1 LDC200CV
Current Control (Constant Current Mode)
Control Range
0 to ±20 mA
Compliance Voltage
>6 V
Resolution
1.0 µA
Accuracy
±20 µA
Noise Without Ripple (10Hz to 10MHz, rms, typ.)
<1.0 µA
Ripple (50/60Hz, rms, typ.)
<0.5 µA
Transients (typ.)
<10 µA
Drift, 24hours
(typ., 0-10Hz, at constant ambient temperature)
<1 µA
Temperature Coefficient
<50 ppm/°C
Current Limit
Setting Range
0 to >20 mA
Resolution
1 µA
Accuracy
± 50 µA
Power Control (Constant Power Mode)
Photo Current Control Range
5 µA to 2 mA
Photo Current Resolution
0.1 µA
Photo Current Accuracy
±2 µA
Analog Modulation Input
Input Resistance
10 kW
Small Signal 3dB Bandwidth, CC Mode
DC to 100 kHz
Modulation Coefficient, CC Mode
2 mA/V ±5%
Modulation Coefficient, CP Mode
0.2 mA/V ±5%
Laser Current Monitor Output
Load Resistance
>10 kW
Transmission Coefficient
500 V/A ±5%
General data
Maximum Power Consumption
20 VA
Weight
<3.1 kg
All technical data are valid at 23 ± 5°C and 45 ±15% humidity

Other manuals for THORLABS LDC200C series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THORLABS LDC200C series and is the answer not in the manual?

THORLABS LDC200C series Specifications

General IconGeneral
BrandTHORLABS
ModelLDC200C series
CategoryController
LanguageEnglish

Related product manuals