38
9. ERROR DISPLAYS
If an error is displayed, follow the procedures shown below.
• If error code remains after trying countermeasures above, contact your local dealer.
Error Display Description/Countermeasure
The Height Alert Function is working.
"6.2 Height Alert Function"
RL-HV setting exceeds the leveling range.
Exceeding the adjustment range.
Turn the power of the RL-HV off, turn on the power back again and readjust.
E-05
Turn the power for the instrument off, and then turn it back on.
E-56
Wireless function error for the RL-HV.
Turn the power for the instrument off, and then turn it back on.
E-65
Internal transmission error for the RL-HV.
Turn the power for the instrument off, and then turn it back on.
E-70’s
Slope function error.
Turn the power for the instrument off, and then turn it back on.
E-80’s
Leveling incomplete.
Turn the power for the instrument off, and then turn it back on in stable
locations.
E-99
Internal memory error for the RL-HV.
Turn the power for the instrument off, and then turn it back on.
Reset tilting to the direction to raise the X+ side.
Reset tilting to the direction to raise the X- side.
Reset tilting to the direction to raise the Y+ side.
Reset tilting to the direction to raise the Y- side.