User Calibration
If a test method is user calibrated the method name is displayed inverse.
Procedure:
•
Prepare a standard of known concentration and use this standard instead of the sample
according to the test procedure.
•
It is recommend to use well known standards which are formulated according to DIN EN,
ASTM or other international norms or to use certified standards which are commercially
available.
•
After measuring this standard solution it is possible to change the displayed results to the
required value.
•
If a method uses a mathematic equation for the calculation of the result, it is only possible
to calibrate the basic tests since all the other tests use the same polynomial.
•
The same applies for some test procedures which use a polynomial from another test
procedure.
Return to factory calibration:
If the user calibration is deleted the factory calibration is automatically activated.
Remarks:
The method ”Fluoride” cannot be calibrated with mode 45 since the test requires a
calibration related to the batch of the liquid reagent (SPADNS)
Recommended range for user calibration
Calibration with basic test 100 Chlorine free
Calibration with basic test 100 Chlorine free
Calibration with 0 and 1 mg/l F through
Mode 40
Calibration with basic test100 Chlorine free
Calibration with basic test 100 Chlorine free