Mode
5
4
100 Chlorine T
0.02-6 mg/l Cl2
0.90 mg/l free Cl2
220
Iron T
0.3 0.7 mg/l Fe
222
Iron PP
0.1 2 mg/l Fe
240
Manganese T
1 2 mg/l Mn
250
Molybdate T
5 15 mg/l Mo
251
Molybdate LR PP
1.52.5 mg/l Mo
257
Nickel T
6 8 mg/l Ni
260
Nitrate LR T
0.50.7 mg/l N
265
Nitrate TT
10 mg/l N
270
Nitrite T
0.2 0.3 mg/l N
300
Ozone (DPD)
Calibration with basic test 100 Chlorine free
290
Oxygen, active
Calibration with basic test 100 Chlorine free
329
pH-Value LR
6.0 6.6
330
pH-Value T
7.6 8.0
332
pH-Value HR
8.6 9.0
320
Phosphate LR T
1 3 mg/l PO
4
321
Phosphate HR T
30 50 mg/l PO
4
340
Potassium
3 mg/l K
350
Silica
0.5 1.5 mg/l SiO
2
212
Sodium hypochlorite
8 %
355
Sulfate T
50 mg/l SO
4
365
Sulfide
0.2 0.4 mg/l S
370
Sulfite
3 4 mg/l SO
3
384
Suspended Solids
operating range
386
Turbidity
operating range
390
Urea
1 2 mg/l CH
4
N
2
O
400
Zinc
0.2 0.4 mg/L Zn
Store user calibration
Perform the required method as described in the manual
using a standard of known concentration instead of the
water sample.
If the test result is displayed press [MODE], [Shift] + [4] [5] keys
and confirm with [ ] key.