8000 SERIES OPERATION MANUAL TRANSMILLE LTD.
TRANSMILLE LTD. PAGE 54
8000 Series DMM Remote Commands
All commands followed by Carriage Return (ASCII 13)
SCPI
Command Format Parameters
Transmit Reading READ?
*TRG
Select AC Voltage Range RANGE:AC:VOLTAGE <value>
RANGE:AC:VOLT <value>
value = 0.01 to 1000
Select DC Voltage Range RANGE:DC:VOLTAGE <value>
RANGE:DC:VOLT <value>
value = 0.1 to 1000
Select AC Current Range RANGE:AC:CURRENT <value>
RANGE:AC:CURR <value>
value = 10
-4
to 30
Select DC Current Range RANGE:DC:CURRENT <value>
RANGE:DC:CURR <value>
value = 10
-8
to 30
Select 4-Wire Resistance RANGE:FRES <value>
RANGE:FRESISTANCE <value>
value=1 to 10
7
Select 2-Wire Resistance RANGE:RES <value>
RANGE:RESISTANCE <value>
value=1 to 10
12
(value may be modified
by SI Prefixes: n, u, m,
k, M, G, T)
General
Command Format Parameters
Use Rear Terminals r
Use Front Terminals f
Set Filter Range F<Filter Range> Filter Range = 1 to 9 for
filter times 250ms – 64s
Set Digits D<Digits> Digits = 4 to 8
Show Status s
Set Null n
Clear Null N
Display Date a
Version Information v
Internal Temperature TEMP?
Example Commands (where <CR> denotes Carriage Return (ASCII 13) :
Required
Output
Range
Command
Output
Value
Command
Separator
Filter
Command
Terminator
(Carriage
Return)
1V DC @ 2s Filter RANGE:DC:VOLT 1 / F4 <CR>
10V AC @ 4s Filter RANGE:AC:VOLT 10 / F5 <CR>
100mA DC @
500ms Filter
RANGE:DC:CUR
R
0.1 / F2 <CR>