V3 / V4 / V5 / V6 / V8
750 50 0045 03 9
5. Slide the measuring insert into the holder and lock
it, using the knob. Take care to position the
eccentric probes correctly.
6. Release the transport safety screw.
7. Check / adjust the floating probe suspension
balancing (§ 10.3).
8. If the instrument does not switch on or in
case of low battery level, proceed to a full
loading of the batteries (plug in the charging
unit to the instrument). Empty batteries are
fully loaded in about 3 hours.
Only use the charging unit supplied with the instrument to load the batteries.
It is not necessary to wait until the battery pack is fully charged. The instrument
becomes operational immediately after having connected the charging unit.
It is not dangerous to leave the charging unit connected. Permanently plugged-in
instruments automatically switch to trickle charge mode.
The batteries are of Lithium-ion type. An excessive room temperature can affect
the battery capacity and therefore the instrument autonomy. It is not
recommended to load the battery in case of room temperature superior to 40C°.
Incorrect loading of the battery can result in a decrease of capacity, overheating,
or even an explosion and cause important damages.
The battery can be loaded and unloaded 300 times before a significant reduction
of their capacity. The number of loading cycles and the autonomy can vary
according to use and operating conditions.
V3
V4
V5
V6
V8
V3, V4, V5, V6
V8