6.5 Probe Holder Change
This function allows to pass from one probe holder to
another while keeping the same origin.
Press the key opposite during 2 seconds to activate the
probe holder change function
> 2 s
Procedure to follow
1. Before taking out the insert probe a reference surface
of centerline and activate the probe holder change
mode. The function symbol is blinking..
> 2 s
2. Mount the probe on the other probe holder.
3. Probe the same surface again, or reference
centreline.
4. Press
Preset
key (the display mentions the latest
measured value.
This procedure can be interrupted at any time by pressing
the
Validate
key.
6.6 Manual Recording of the Probe Constant
The probe constant can be introduced manually. To do
so, press the opposite function key during 2 seconds and
enter the requested value with the numeric keypad.
Confirm by pressing the
Function
key.
The registered value is taken into account during
measurements. Each new constant acquisition replaces
the old one.
> 2 s
The constant is an essential element during bidirectional measurements. Great
caution should be applied for the interpretation of the measurements results when
this value is entered manually.