8. REMOTE CONTROL OF THE INSTRUMENT
The instrument can be remotely-controlled by ASCII commands via the RS232 interface.
- The RS232 interface shall be configured to OPTO RS (see § 9).
- Up to 64 commands can be executed simultaneously.
- When a set of commands is executed, it is not possible to access keyboard functions.
- After the first error, all following commands are cancelled.
- When an error occurs, no new command is accepted except: "ERR?", "!ERR" et "RST".
8.1 Command attributes
Some commands can only be executed if the instrument is in a given configuration
(command attributes). If a command is sent while the attribute is not respected, an error will
be generated.
List of command attributes:
1. ATTR_REF => The reference must habe been passed
2. ATTR_CONST => The probe constant must have been taken.
3. ATTR_MOTOR => The motorised mode must be switched on.