V3 / V4 / V5 / V6 / V8
750 50 0045 03 29
When the shrinking factor is different from 1, the following
symbol appears at the top of the main display.
6.9 Temperature Compensation
When the ambient temperature differs from 20°C, it is possible to compensate the
measurement according to the actual room temperature and the thermal expansion
coefficient of the part. The displayed values are then recalculated to the standard
temperature 20 °C.
To activate the temperature compensation, press the
following button during 2 seconds
> 2 s
Enter the real room temperature (value between 10 and
40, unit [°C]) and confirm by pressing
.
Enter the value of thermal expansion coefficient of the
part (value between 0 and 200, unit [10
-6
K
-1
]) and confirm
by pressing
.
If the entered temperature is different from 20 °C, the
function indicator will display "T".