T3-OPX installation
Page | 76
13.1 Controls for T1030 (webpath board)
Typical commands to use, these can be run through the webpath accessed through putty or directly
through teraterm (USB mini connection):
Function
Command
Comment
Move the unit up or down
dcseekmm <value in mm 0-100>
0 is the highest position, 100 is
the lowest position.
Calibrate actuators
dcz
Unit is moved to the top and
calibrated (tilt etc)
Calibrate the table position
height and x nad y tilt, to create
the base settings.
calall
Read the power state of the
mpca
Mpcap
It will return 1 if the power is on
Push the virtual power button of
the MPCA
Mpcap (2 or 0)
2 the button is pressed, 0 the
button is released.
To switch it on
>mpcap 2
wait 2 seconds then enter
>mpcap 0
Power control of mpca, fans etc
V33off <0,1 or blank>
Blank reads the value
0 power is provided
1 power is cut
Read the weight on of the
individual load sensor, there are
4 in total
Lgram <1,2,3 or 4>
Read the total weight on the
load sensors
Lgramt
No parameters