EasyManua.ls Logo

UNI-T UTG900E Series - Chapter 4 Advanced Applications; 4.1 Generate Modulation Waveform

UNI-T UTG900E Series
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Manual UTG900E Series
Instruments.uni-trend.com
14 / 41
<voltage> means voltage and its unit is the specified unit of current channel.
<n>: Channel No, n value 1, 2.
Return Format
The query returns the high signal output value of specified channel, using scientific notation to return.
Example
:CHANnel1:BASE:HIGH 2
Set the high signal output value of channel 1 to 2V
:CHANnel1:BASE:HIGH?
The query returns 2e+0
:CHANnel<n>:BASE:LOW
Command Format
:CHANnel<n>:BASE:LOW { <voltage>}
:CHANnel<n>:BASE:LOW?
Function
Set the low signal output value of specified channel.
<voltage> means voltage and its unit is the specified unit of current channel.
<n>: Channel No., n value 1, 2.
Return Format
The query returns the low signal output value of specified channel, using scientific notation to return.
Example
:CHANnel1:BASE:LOW 2
Set the low signal output value of channel 1 to 2V
:CHANnel1:BASE:LOW?
The query returns 2e+0
:CHANnel<n>:BASE:DUTY
Command Format
:CHANnel<n>:BASE:DUTY { <duty>}
:CHANnel<n>:BASE:DUTY?
Function
Set the duty ratio of signal output in specified channel.
<duty> means the duty ratio, in “% unit, range of 0~100.
<n>: Channel No., n value 1, 2, 3, 4.
Return Format
The query returns the duty ratio of signal output in specified channel.
Example
:CHANnel1:BASE:DUTY 20
Set the duty ratio of signal output in channel 1 to 20%
:CHANnel1:BASE:DUTY?
The query returns 20
:CHANnel<n>:RAMP:SYMMetry
Command Format
:CHANnel<n>:RAMP:SYMMetry { < symmetry >}
:CHANnel<n>:RAMP:SYMMetry?
Function