Programming Manual UTG900E Series
Instruments.uni-trend.com
5 / 41
SCPI Details
IEEE488.2 Common Commands
*IDN?
Command Format
*IDN?
Function
Query the manufacturer name, signal source model, product serial number and software versioning.
Return Format
The query returns the manufacturer name, signal source model, product serial number and the
software versioning separated by dot.
Note: The returning model should be same as the nameplate.
Example
UNI-T Technologies, UTG900, 000000001, 00.00.01
*RST
Command Format
*RST
Function
Restore factory settings, clear all error messages and send/receive queue buffers.
SYSTem Commands
It is used for the most basic operation of signal source, mainly including the operation of full keyboard lock
and the system data setting.
:SYSTem:LOCK
Command Format
:SYSTem:LOCK {{1 | ON} | {0 | OFF}}
:SYSTem:LOCK?
Function
Lock or unlock the full keyboard keys and touch input.
Return Format
The query returns the full keyboard status, and 0 in UNLOCK, 1 in LOCK.
Example
:SYSTem:LOCK ON Full keyboard LOCK
:SYSTem:LOCK OFF Full keyboard UNLOCK
:SYSTem:LOCK? Query returns 1, LOCK
:SYSTem:CONFigure
Command Format
:SYSTem:CONFigure <file>
:SYSTem:CONFigure?
Function
Read and write the configuration file, sending commands first, and send configuration file data to the
signal source.

Table of Contents

Other manuals for UNI-T UTG900E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UTG900E Series and is the answer not in the manual?

UNI-T UTG900E Series Specifications

General IconGeneral
BrandUNI-T
ModelUTG900E Series
CategoryPortable Generator
LanguageEnglish

Related product manuals