EasyManua.ls Logo

UNI-T UTG900E Series

UNI-T UTG900E Series
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Manual UTG900E Series
Instruments.uni-trend.com
20 / 41
:CHANnel1:FSK:HOPP:FREQ?
The query returns 2e+3
:CHANnel<n>:SWEep:FREQuency:STARt
Command Format
:CHANnel<n>:SWEep:FREQuency:STARt <freq>
:CHANnel<n>:SWEep:FREQuency:STARt?
Function
Set the start frequency of sweep in specified channel.
< freq > means frequency, in “Hz” unit.
<n>: Channel No., n value 1, 2.
Return Format
The query returns the start frequency of sweep in specified channel, using scientific notation to return
the data.
Example
:CHANnel1:SWE:FREQ:STAR 2000
Set the start frequency of sweep in channel 1 to 2KHz
:CHANnel1:SWE:FREQ:STAR?
The query returns 2e+3
:CHANnel<n>:SWEep:FREQuency:STOP
Command Format
:CHANnel<n>:SWEep:FREQuency:STOP <freq>
:CHANnel<n>:SWEep:FREQuency:STOP?
Function
Set the stop frequency of sweep in specified channel.
< freq > means frequency, in “Hz” unit.
<n>: Channel No., n value 1, 2.
Return Format
The query returns the stop frequency of sweep in specified channel, using scientific notation to return
the data.
Example
:CHANnel1:SWE:FREQ:STOP 2000
Set the stop frequency of sweep in channel 1 to 2KHz
:CHANnel1:SWE:FREQ:STOP?
The query returns 2e+3
:CHANnel<n>:SWEep:TIMe
Command Format
:CHANnel<n>:SWEEP:TIMe { <time>}
:CHANnel<n>:SWEEP:TIMe?
Function
Set the sweep time of specified channel.
< time > means time, in “S unit, range of 1ms ~ 500s.
<n>: Channel No., n value 1, 2.
Return Format
The query returns the sweep time of specified channel, using scientific notation to return the data.

Table of Contents