EasyManua.ls Logo

UNI-T UTG900E Series

UNI-T UTG900E Series
41 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Manual UTG900E Series
Instruments.uni-trend.com
7 / 41
:SYSTem:BEEP ON Turn buzzer on
:SYSTem:BEEP? The query returns 1
:SYSTem:NUMBer:FORMat
Command Format
:SYSTem:NUMBer:FORMat {COMMa|SPACe|NONe}
:SYSTem:NUMBer:FORMat?
Function
Control the separator of system number format.
Return Format
The query returns the separator of system number format.
Example
:SYSTem:NUMBer:FORMat NONe
Set the system number format to be none
:SYSTem:NUMBer:FORMat?
The query returns NONe
:SYSTem:BRIGhtness
Command Format
:SYSTem:BRIGhtness { 10|30|50|70|90|100}
:SYSTem:BRIGhtness?
Function
Control the brightness of system backlight.
Return Format
The query returns the brightness of system backlight.
Example
:SYSTem:BRIGhtness 30
Set the brightness of system backlight to 30%
:SYSTem:BRIGhtness?
The query returns 30
:SYSTem:SLEEP:TIMe
Command Format
:SYSTem:SLEEP:TIMe { CLOSe | 1 |5 | 15 |30 |60}
:SYSTem:SLEEP:TIMe?
Function:
Control the sleep time of system, in “Minute” unit.
Return Format
The query returns the sleep time.
Example
:SYSTem:SLEEP:TIMe 5 MIN
Set the system to auto-sleep in 1 minute
:SYSTem:SLEEP:TIMe?
The query returns 1
:SYSTem:CYMometer
Command Format
:SYSTem:CYMometer {{1 | ON} | {0 | OFF}}

Table of Contents

Other manuals for UNI-T UTG900E Series

Related product manuals