4.CALIBRATION
21
Actual Load Calibration
This approach provides calibration by applying an actual load to the strain gauge sensor
and inputting the actual load value. This calibration is without little errors and more
correct.
4-1. Equivalent Input Calibration Procedure
The equivalent input calibration uses the following procedure :
Release the Calibration LOCK that inhibits calibration.
Set the minimum value of the digital change desired
omissible if no desired value is present
Set the rated output vale and zero point of the strain
gauge sensor.
Set the calibration of ON to prevent misoperation.
F340A
Strain gauge sensor
Indicated value
Actual
load
Releasing Calibration LOCK
Setting Scale Division
Equivalent Input Calibration
Inhibit Calibration
Calibration Lock