10.D/A CONVERTER
73
10-1-1. Obtaining Voltage Output Signal
Use + and - terminals of F340A connecting to them an external equipment (with load
resistance of 2kΩ or more).
10-1-2. Setting of D/A Zero and Gain
With the D/A converter of the F340A, an analog output is obtained by setting the
indicated value to output 0V (D/A zero set value) and the indicated value to output 10V
(D/A full scale set value). Respective set values are inputted by the D/A zero and full
scale setting functions.
+
-
+
-
External equipment
Load resistance: 2kΩ or more
SIG
GND
← Inside Outside →
F340A
CAUTION
• Do not apply external voltages. Breakage will result.
• Do not short-circuit the voltage output. Doing so will cause a failure.
Also, connecting a capacity load may cause oscillation.
Voltage
(V)
10
-10
Indicated value
0
Resolution 1/6000
D/A full scale set valueD/A zero set value