15.SPECIFICATIONS
88
15. SPECIFICATIONS
15-1. Analog Section
Bride Voltage DC 10V ± 10
DC 2.5V ± 10
Output current of ax 30mA
Changable by setting key Pad
Signal input range 3.0 to +3.0mV/V
Equivalent input calibration range +0.5 +3.0mV/V
Equivalent input calibration error 0.1FS0.5mV/V input
Zero adjustment range 0 ~± 2.0mV/V
Gain adjustment range +0.5 +3.0mV/V
Analog input signal sensitivity 1μV/count
Accuracy Non-linearity:< 0.02%FSof 3mV/V input
Zero drift :< 0.5μV/
Gain drift :< 25ppm/
A/D converter 100 times /sec. Resolution16 bits binary
Analog filter 4Hz10Hz100HzInitial3Hz
Changable by setting key pad
Peak hold function high-speed analog hold system
Operation response speedApprox. 1kHz
Sin wave : 3mV/V input
Analog Filter.3kHz
Accuracy 0.1%FS
Reset time:< 50μS