5.SETTING OF FUNCTIONS
49
5-13. LOCK
The Setting value LOCK function inhibits changes to setting to prevent changes to set
values or calibrated values through misoperation.
1) Select setting mode 3.
2) Select lock.
Use and key to set the lock,
then use key to validate the setting.
To return to the indicated value display, press key.
Setting of LOCK
1 : LOCK
0 : LOCK
released
Setting Value LOCK
1 : LOCK
0 : LOCK
released
Calbration LOCK
Zero key operation
1 :Invalid
0 :Val id
Hold key
1 :Invalid
0 :Valid
operation
For setting items locked by using set value LOCK and calibration
value LOCK , see the set value list on page 17.
The Digital Zero function by using the key on the front
panel is not effective when the Setting value LOCK (calibrated
value)is released(nor is it effective from the rear panel DZ).
On completion of calibration set the calibration value LOCK.