4.CALIBRATION
28
・Setting of Scale Division(omissible if no change is needed)
* When you set the actual load calibration, the change that digit to place a decimal point
is set.
・Zero Calibration
1) Select setting mode 3.
2) Set the scale division.
Use and keys to set the scale division,
then use key to validate the setting.
To return to the indicated value display, press key.
Setting of Scale Division
Scale Division(001to 100)
1)Check that the sensor is unloaded, then press keys in
this order.
If the indicated value following the display CAL2E is
zero, the zero calibration is terminated.