EasyManua.ls Logo

UseTECH FlowTimer+ - Page 44

UseTECH FlowTimer+
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting, maintenance
01/21 FlowTimer+ Set 44
If the leakage is still not indicated, temporarily close the KFR
valve on the water meter (see also Figure 10.2.2).
Fig. 10.2.2: KFR valve water meter
Now check again according to point 8.3. "Verification of leakage".
After the check, please open the KFR valve again
immediately to prevent the system from running empty via
the simulated leak.
If a leaking system is now indicated, you should have the shut-off
valve of the FlowTimer+ cleaned.
Residues of sealing tapes or chips from connectors may have
entered the valve during installation.
WARNING
The system must first be disconnected from the power
supply, shut off from the water supply and
depressurized.

Other manuals for UseTECH FlowTimer+