User’s Guide ______________________________________________________________________
54 ___________________________________________________________________ M010056EN-I
This command outputs the current value of each channel and the
corresponding control signal of the digital-to-analog converter.
Example:
>itest 8 12
8.00000 403 12.00000 7DF
>itest
7.00150 30A 11.35429 73E
>
The set current values remain valid until you issue the command ITEST
without readings or reset the transmitter. With this command the desired
outputs of the transmitter are shown.
SEND Outputting Measurement Values
Syntax: SEND<cr>
This command outputs the measured values in one point.
R Activating Continuous Output
Syntax: R<cr>
With the command R the transmitter outputs measured values
continuously.
S Stopping Continuous Output
Syntax: S<cr>
The continuous outputting is stopped with the command S. If outputting is
active, this command is not echoed.
INTV Setting Output Interval
Syntax: INTV [n xxx]<cr>
NOTE
When outputting low currents from Ch 1, remember to remove RS232C-
cable while reading the current output, because of the increased current
consumption for using RS port.
where