EasyManua.ls Logo

Vaisala HUMICAP HMT360 SERIES - Accuracy of Calculated Variables; Accuracy of Dewpoint Temperature °C; Accuracy of Mixing Ratio G;Kg

Vaisala HUMICAP HMT360 SERIES
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 ____________________________________________________________ Technical Data
VAISALA_______________________________________________________________________ 81
Accuracy of Calculated Variables
Accuracy of the calculated variables depend on the calibration accuracy of
the humidity and temperature sensors; here the accuracy are given for ±2
%RH and ±0.2 °C.
Accuracy of Dewpoint Temperature °C
Accuracy of Mixing Ratio g/kg
(Ambient pressure 1013 mbar)
Relative humidity
Temp.102030405060708090100
-40 1.86 1.03 0.76 0.63 0.55 0.5 0.46 0.43
-20 2.18 1.19 0.88 0.72 0.62 0.56 0.51 0.48
0 2.51 1.37 1 0.81 0.7 0.63 0.57 0.53 0.5 0.48
20 2.87 1.56 1.13 0.92 0.79 0.7 0.64 0.59 0.55 0.53
40 3.24 1.76 1.27 1.03 0.88 0.78 0.71 0.65 0.61 0.58
60 3.6 1.96 1.42 1.14 0.97 0.86 0.78 0.72 0.67 0.64
80 4.01 2.18 1.58 1.27 1.08 0.95 0.86 0.79 0.74 0.7
100 4.42 2.41 1.74 1.4 1.19 1.05 0.95 0.87 0.81 0.76
120 4.86 2.66 1.92 1.54 1.31 1.16 1.04 0.96 0.89 0.84
140 5.31 2.91 2.1 1.69 1.44 1.26 1.14 1.05 0.97 0.91
160 5.8 3.18 2.3 1.85 1.57 1.38 1.24 1.14 1.06 0.99
Relative humidity
Temp.102030405060708090100
-40 0.003 0.003 0.003 0.003 0.003 0.004 0.004 0.004
-20 0.017 0.018 0.019 0.021 0.022 0.023 0.025 0.026
0 0.08 0.09 0.09 0.1 0.1 0.11 0.11 0.12 0.13 0.13
20 0.31 0.33 0.35 0.37 0.39 0.41 0.43 0.45 0.47 0.49
40 0.97 1.03 1.1 1.17 1.24 1.31 1.38 1.46 1.54 1.62
60 2.68 2.91 3.16 3.43 3.72 4.04 4.38 4.75 5.15 5.58
80 6.73 7.73 8.92 10.34 12.05 14.14 16.71 19.92 24.01 29.29
100 16.26 21.34 28.89 40.75 60.86 98.85 183.66 438.56
120 40.83 74.66 172.36

Table of Contents