Register number Address Register
description
Data format Bit order
15 000E
hex
Gas level alarms
for present values
16-bit field 0 = Methane
1 = Acetylene
2 = Ethylene
3 = Ethane
4 = Carbon
monoxide
5 = Carbon
dioxide
6 = TDCG
7 = Hydrogen
8 = Moisture in oil
9 = Relative
moisture in oil
11 = TGP above
maximum limit
12 = TGP below
minimum limit
15 = Any alarm
active
16 000F
hex
Gas level alarms
for 24 hour
average ROC
values
16-bit field 0 = Methane
1 = Acetylene
2 = Ethylene
3 = Ethane
4 = Carbon
monoxide
5 = Carbon
dioxide
6 = TDCG
7 = Hydrogen
15 = Any alarm
active