3.7 Configuring autocalibration
Figure 29 Autocalibration page
Starting with software version 1.11.0, the DGA monitor performs autocalibration when needed
and when environmental conditions are suitable. Vaisala recommends leaving this feature
enabled, but Admin user can disable it from the Autocalibration page.
Even if the automatically started autocalibration is disabled, the initial autocalibration will still
run after the DGA monitor has been initialized. Autocalibration can also be manually started.
1. Connect to the user interface as the Admin user.
2. In the user interface, select Control > Autocalibration.
3. Configure the setting with the switch:
• Disable automatically started autocalibration by moving the switch to the left.
• Enable automatically started autocalibration by setting the switch to position Start
automatically.
More information
‣
Autocalibration (page 28)
‣
Connecting to user interface (page 30)