When the test mode is enabled, normal measurement operation of the DGA
monitor continues in the background. Only the data available through the digital
output protocols is aected. When the mode ends, real measurement data and
alert status are again available through the protocols.
More information
Using protocol test mode (page 62)
2.16 Autocalibration
The autocalibration feature of the DGA monitor improves measurement performance by
adapting the measurement to the individual oil conditions of the transformer. Autocalibration
is implemented as special type of measurement cycles that vary the parameters of the vacuum
gas extraction. The process typically completes in two days, but may take longer if ambient
temperature is high. After autocalibration is completed, the measurement algorithm is
adjusted according to the analysis results, and normal measurement cycles are automatically
resumed.
DGA monitor makes use of autocalibration in the following ways:
Table 9 Types of autocalibration
Autocalibration type Description Notes
Initial autocalibration Initial autocalibration runs
automatically after a DGA
monitor has been initialized
and measurement started.
Initial autocalibration cannot be
skipped, and it must complete
successfully for DGA monitor to
operate normally.
Measurement results generated
during the initial
autocalibration may not be
accurate. Gas level alerts are
not indicated during initial
autocalibration via any means,
including relays.
Manually started
autocalibration
One-time autocalibration that
is started manually by the
Admin user.
Recommended after:
Transformer degassing or oil
change
DGA monitor software
update
DGA monitor repair if
components of measurement
electronics have been
replaced
OPT100 User Guide M211858EN-U