4. Select Apply to take the new alert limit settings into use.
Changing gas alert settings will not immediately activate any new alerts. Gas
alert status is updated only when the next measurement cycle results set is
ready.
More information
Connecting to user interface (page 30)
3.9 Muting gas level alerts
Muting gas level alerts temporarily disables any indications that any gas level alerts are active:
Alert indicator in the user interface is set to non-alerting state (unless any device alerts
are active)
If one or more relays have been activated by gas level alerts, they are inactivated
Status LEDs do not indicate that a gas level alert is active
Output protocols do not communicate any gas level alerts as active (but will set
appropriate status bits/flags to signal that gas level alarms are muted)
Admin user can mute gas level alerts by selecting Mute gas alerts on the Alerts page.
Muting of gas level alerts continues until any one of the following occurs:
Severity of any muted gas level alert rises from caution to alarm
Any new gas level alert is activated
Admin user selects Unmute gas alerts on the Alerts page
OPT100 User Guide M211858EN-U