Alerts page shows the current alert status of the DGA monitor.
1. View number of currently active gas level alerts and device alerts by severity.
2. See Events section for more information on the individual alerts. By default, only active
alerts are listed.
a. To control what is shown, select the desired options for Status, Type, and Severity.
b. To temporarily suppress gas level alerts, press Mute gas alerts.
c. To view detailed information on a listed event, select Details. Details depend on the
type of alert being viewed. Details for device alerts include information on resolving
the alert.
You can also configure custom gas level alerts.
Figure 19 Event details for a gas caution alert
More information
‣
Configuring gas level alerts (page 53)
‣
Muting gas level alerts (page 54)
Chapter 3 – User interface