EasyManua.ls Logo

Vaisala PTU300 - Filt; Relative Humidity (RH) and Temperature (T) Filtering; Table 31 Filtering Levels for Relative Humidity (RH) and Temperature (T)

Vaisala PTU300
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 ________________________________________________________________ Operation
VAISALA______________________________________________________________________ 111
Relative Humidity (RH) and Temperature (T) Filtering
Table 31 Filtering Levels for Relative Humidity (RH) and
Temperature (T)
Setting Filtering Level
OFF No filtering
ON (default) Standard = short filtering (approximately 15 s moving
average)
EXTENDED Extended filtering (default: approximately 1 min
average)
Use display/keypad to set the filtering level.
1. Press any of the arrow buttons to open the Main Menu.
2. Select Measuring and press the right arrow button.
3. Select Filtering and press the CHANGE key.
4. Select Off/Standard/Extended and press the SELECT key.
5. Press the EXIT key to return to the basic display.
FILT
Use the serial line command FILT [xxx] to set the filtering level.
FILT [xxx]<cr>
where
xxx = OFF, ON, or EXT (default = OFF)

Table of Contents

Related product manuals