EasyManua.ls Logo

Vaisala PTU300 - Relative Humidity Adjustment after Sensor Change; Using Display;Keypad; Using Serial Line; Fcrh

Vaisala PTU300
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7 ___________________________________________________Calibration and adjustment
VAISALA______________________________________________________________________ 173
Relative Humidity Adjustment After Sensor
Change
Using Display/Keypad
When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 170 on page but select Adj. for
new RH sensor (instead of 1-point/ 2-point adjustment).
Using Serial Line
After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.
FCRH
Example:
>fcrh
RH : 1.82 1. ref ? 0
Press any key when ready...
RH : 74.22 2. ref ? 75
OK
>
The OK indicates that the calibration has succeeded.

Table of Contents