Chapter 5 ________________________________________________________________ MODBUS
VAISALA______________________________________________________________________ 149
Enabling Serial MODBUS
In addition to enabling the MODBUS protocol in the device, MODBUS
over RS-232 or RS-485 needs the following software settings: serial bit
rate, parity, number of stop bits, and MODBUS device address.
Using Display/Keypad (Optional)
1. Open the Main Menu by pressing any of the ▼▲◄► arrow
buttons.
2. Navigate to Interfaces Serial Interface.
3. On the Serial Interface settings screen:
- Enable the MODBUS protocol.
- Change the baud rate and parity if necessary.
- Set the MODBUS device address.
1101-033
Figure 67 Serial Interface Settings
4. After changing the settings, press the EXIT button. The MODBUS
configuration is now complete, as configuration changes made with
the display and keypad are effective immediately.
Using Serial Line
1. Open the transmitter cover.
2. Connect power supply wiring (if not using the power supply
module) and power up the transmitter.
3. Connect the USB service cable between a computer and the service
port of the transmitter.
4. Start the Vaisala USB Instrument Finder program (which has been
installed on the computer along with the USB service cable driver),
and check the COM port that the cable is using.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala PTU300 and is the answer not in the manual?

Vaisala PTU300 Specifications

General IconGeneral
Measurement ParametersTemperature, Humidity, Pressure
Temperature Range-40 to +60 °C
Humidity Range0 to 100 %RH
Operating Temperature-40 to +60 °C
Operating Humidity0 to 100 %RH
Protection ClassIP65
Power Supply10 to 30 VDC
Communication ProtocolsModbus
Measurement accuracy of humidity±1.5 %RH
Measurement accuracy of temperature±0.2 °C
Output SignalRS-485