USER'S GUIDE____________________________________________________________________
94 __________________________________________________________________ M210796EN-F
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
In the tables below, the bold text in the brackets is the default setting.
Table 19 Measurement Commands
Command Description
R Start the continuous outputting
S Stop the continuous outputting
INTV [0 ... 255 S/MIN/H] Set the continuous output interval
(for RUN mode)
SEND [0 ... 255] Output the reading once
SEND D Outputting the reading with the raw data
SCOM Define a user-specific SEND command for one
message output
SMODE [STOP/SEND/RUN/
POLL/MODBUS]
Set the serial interface mode
SDELAY View or set user port (RS-232 or RS-485)
answer minimum delay
SERI [baud p d s] User Port settings (Default: 4800 E 7 1) baud:
300 ... 115200
ADDR [0 ... 255] Set the transmitter address (for POLL and
MODBUS modes)
NET View or set networking parameters for LAN and
WLAN interfaces
WLAN View or set wireless network parameters for
WLAN interface
OPEN [0 ... 255] Open a temporary connection to a POLL mode
device
CLOSE Close the temporary connection (Back to POLL
mode)
Table 20 Formatting Commands
Command Description
FORM Set the output format of SEND and R
commands
TIME Set the time
DATE Set the date
FTIME [ON/OFF] Add time to SEND and R outputs
FDATE [ON/OFF] Add date to SEND and R outputs
FST [ON/OFF] Add the state of probe heating and chemical
purge in connection with SEND and R
commands
SCOM Assign a new command name that works like
the SEND command
UNIT Select the metric or non-metric output units