EasyManua.ls Logo

Vaisala WINDCAP WMT52 - Table 10 Communication Problems

Vaisala WINDCAP WMT52
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User’s Guide ______________________________________________________________________
110 __________________________________________________________________M210925EN-B
Table 10 Communication Problems
Problem Interpretation Action
No response to any commands. Wrong wiring or operation
voltage not connected.
Baud rate/start bits/parity/stop bit
settings do not match between
the device and the host.
Check the wiring and operation
voltage, see Chapter 5, Wiring
and Power Management, on
page 39.
Connect the service cable, use
the communication settings
19200,8 N,1. Check the serial
port settings of the device with
Configuration Tool or with
terminal program. Use command
aXU! (SDI-12) or aXU<cr><lf>
(ASCII/NMEA). Change the
values if needed. A software/
hardware reset is needed to
validate the changes.
When you have no service
cable, try typing address query
commands ?! and ?<cr><lf> with
different serial settings in
terminal program. When the
communication parameters
match, the device responds with
its address. The settings can
now be changed using aXU!
(SDI-12) or aXU<cr><lf> (ASCII/
NMEA) commands. A software/
hardware reset is needed to
validate the changes.
Incorrect <cr><lf> settings in the
terminal program.
In ASCII and NMEA protocols
both <cr> and <lf> are required
after each command. Check that
your terminal program sends
both when pressing enter.
Connection works but data
messages not available.
Wrong device address in a SDI-
12 command or a mistyped SDI-
12 command (in SDI-12 a
mistyped command is in no way
responded).
A command mistyped in ASCII/
NMEA mode while error
messaging/text messages is
disabled (aSU,S=N).
Request the device address with
?! command and then retype the
command now with the correct
address. Check the data query
commands from Chapter 7,
Getting the Data Messages, on
page 61.
Enable the error messaging
using the Vaisala Configuration
Tool or any terminal by setting
aSU,S=Y, then try the command
again.

Table of Contents

Related product manuals