Chapter 7 __________________________________________________  Getting the Data Messages
VAISALA_______________________________________________________________________ 79
Examples of aM, aC and aD 
Commands
The device address is 0 in all examples.
Example 1:
Start a wind measurement and request the data (all six wind parameters 
are enabled in the message):
0M1!00036<cr><lf> (measurement ready in 3 seconds and 6 
parameters available)
0<cr><lf> (measurement completed)
0D0!0+339+018+030+0.1+0.1+0.1<cr><lf>
<cr><lf> = Response terminator
NOTE
aD0 command can also be used to break the measurement in progress 
started with commands aM, aMC, aC or aCC.
NOTE
In SDI-12 v1.3 Continuous measurement mode (aXU,M=R) the 
sensor makes measurements at configurable update intervals. The aD 
command following the aM, aMC, aC or aCC command always 
returns the latest updated data. Thus in aXU,M=R mode issuing 
consecutive aD commands may result in different data strings if the 
value(s) happen to be updated between the commands.
NOTE
The parameter order in messages is as follows:
Wind (M1): Dn Dm Dx Sn Sm Sx
Supv (M5): Th Vh Vs Vr Id
Comp (M): Wind Supv (parameters in above order)
The parameters are in the same order as in the Setting Field tables in 
chapter Sensor and Data Message Settings on page 95. The order of the 
parameters is fixed, but you can exclude any parameter from the list 
when configuring the sensor.