SECTION
1
Force 2 Service Manual 1-1
1Service Warnings and Cautions
Danger
Explosion Hazard Do not install the Force 2 generator in the presence of
flammable anesthetics, gases, liquids, or objects.
For continued protection against fire hazard, replace the fuse with those having
the same type and rating as the original fuse.
Warning
Electric Shock Hazard Always unplug the generator before cleaning.
Electric Shock Hazard Do not connect a wet power cord assembly into the
generator or into the wall outlet.
Electric Shock Hazard Disconnect the power before replacing parts.
Electric Shock Hazard Do not touch any exposed wiring or conductive surface
while the generator is disassembled and energized. Never wear a grounding
strap when working on an energized generator.
Keep the handset electrode away from personnel. Use the accessory holster to
hold the handset safely.
Use the ACCESSORY output receptacle for connecting either a handswitch (three-
pin) or footswitch (one-pin) accessory, but not both at the same time. Connecting
more than one accessory to the ACCESSORY output receptacle activates both
accessories simultaneously.
Electric Shock Hazard The power FETs and other components on the supply
heatsink are at line potential. Use extreme caution when probing this circuitry.
Take appropriate precautions (such as use of isolated tools and equipment; use
of the “one hand rule,” etc.) when taking measurements or troubleshooting the
generator.