SECTION
6
Force 2 Service Manual 6-1
6Maintenance Procedures
Routine Maintenance and Inspections
Valleylab recommends that qualified service personnel inspect the Force 2
generator every six months. Check the power cord assembly periodically
for damaged insulation or connectors.
This service manual describes the recommended maintenance, inspection,
testing, and calibration procedures. Return the generator to Valleylab for
major repairs. Repair parts and information are available from Valleylab.
Refer to Section 8, Service Parts List, for additional information.
Cleaning Instructions
Clean the Force 2 generator using standard hospital procedures. Use a
mild detergent and damp cloth to clean the cover, control panel,
footswitch, and cord. Do not allow fluids to enter the chassis. Do not use
caustic, corrosive, or abrasive cleaning materials. The generator and all
components must be completely dry before use.
Do not sterilize the Force 2 generator.
Warning
Electric Shock Hazard Do not install a wet power cord assembly into the generator
or into the wall outlet.