Output Power vs. Resistance Graphs
Technical Specifications
ForceTriad User’s Guide 9-19
Fulgurate
Output power versus impedance
for Fulgurate power
Output power versus power setting
for Fulgurate power
Peak voltage versus power setting
for Fulgurate power
0
20
40
60
80
100
120
140
0 1000 2000 3000 4000 5000
50%
100%
Load Impedance (ohms)
Output power (watts)
0
20
40
60
80
100
120
0 20 40 60 80 100 120
Fulgurate Output
Power vs . Setting
@500ohm
Output power (watts)
Power Setting
0
500
1000
1500
2000
2500
3000
3500
0 20 40 60 80 100 120
Open Circuit
Fulgurate Peak
Voltage vs. Power
Setting
Power Setting
Peak Voltage