Caution
Do not use any sharp object to operate the touch screen of the product; otherwise, the screen
may be damaged.
Never apply excessive force to the connector; otherwise, the product may be damaged.
5.1. Preparation and Inspection
Warning
To power supply problems, a power supply with medical certification specified in the manual is
required. When the power supply needs to be connected, only the type and manufacturer as
stated in the packing list is allowed. Otherwise, operator or patient injury may occur.
To avoid the risk of electric shock, only the appropriate approved medical electrical equipment are
allowed to be connected to the product.
The product can only be connected to a medical display screen that has been qualified as per
IEC 60601-1.
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment may result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation
.
Do not use the product in case of any damage to it or any unacceptable item in the functional
inspection. Otherwise, operator/patient injury may occur.
1. Prepare Digital Video Monitor, power cable, power adapter, cables and device used together.
Carefully check the product and all the components for damage (and wear). Never use the product in
case of any damage.
2. Prepare and inspect the endoscope and other accompanying equipment according to their user
manual.
5.2. Installing the Digital Video Monitor
Warning
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth. Electric shock and/or fire may be caused if the product is not properly grounded.
All the system components shall be powered off before connecting them. Only appropriate cables
can be used for the product; Otherwise, the product may be damaged or malfunction.
Before connecting the endoscope to the Digital Video Monitor , ensure that the Digital Video
Monitor is completely dry. Otherwise, it may cause electric shock or equipment damage.
The equipment is installed on a VESA 75 standard rack. It is recommended to choose the screws