7. Cleaning and Disinfection
The product should be cleaned and disinfected before and after each use. It is recommended to clean
and disinfect the product according to the following instructions before and after use. Any deviations
from the instructions shall be assessed properly to determine its effectiveness and potential adverse
consequences to ensure that it continues to achieve its intended purpose.
Warning
Disconnect the Digital Video Monitor from any mains power supply, remove any accessories and
make sure the Digital Video Monitor is completely turned off before cleaning and disinfection.
You should take the following steps to clean and disinfect the product according to good medical
practices:
7.1. Cleaning
1. Recommended detergent: enzymatic, mild pH of 7–9, and low foam (Enzol or equivalent).
2. Prepare the cleaning solution by using the standard enzymatic detergent prepared according to
the manufacturer's recommendations.
3. Dip the sterile gauze into the enzyme solution and ensure that the gauze is moist without
dripping.
4. Use the wet gauze to thoroughly clean the buttons and housing of the product. Well protect the
product from being wet to avoid damage to its internal electronic components.
4. Use a sterile soft brush dipped in enzyme solution to clean the buttons until all dirt is removed.
5. Wait 10 minutes (or the duration recommended by the detergent manufacturer) for the enzyme to
be activated.