Statement
The product is designed compatible with the high-frequency surgical equipment, i.e. it is allowed to
keep working or standby near the high-frequency surgical equipment.
In case of interrupted AC input voltage, the product will shut down, and once the power supply is
restored, it can be manually restored by the operator. The degradation in such case is acceptable
because it will neither cause unacceptable risks nor impair basic safety or essential performance.
The degradation due to electrostatic discharge or electrical fast transient/burst is acceptable because
it will neither cause unacceptable risks nor impair basic safety or essential performance.
The vertical-line flashing on the screen can automatically restore to the previous state.
EMI compliance
Table 1. Emission
Electromagnetic Environment
Professional medical environment
Professional medical environment
Voltage fluctuation
and flicker
Professional medical environment
Caution: The emission characteristics of the product make it suitable for use in industrial areas and
hospitals (CISPR 11 Class A). If being used in the residential environment (CISPR 11 Class B
required usually), the product may not provide adequate protection for radio frequency
communication services. The user may be required to take mitigation measures, for example,
relocating or redirecting the product.
EMS compliance
Table 2. Enclosure port
Professional medical environment
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated,
radio-frequency,
electromagnetic field
3 V/m
80 MHz–2.7 GHz
80% AM at 1 kHz
Fields near the
radio-frequency
wireless
communication
devices
Rated power
frequency magnetic
field