EasyManua.ls Logo

Vathin DVM-D1 - Appendix 1. Electromagnetic Compatibility

Vathin DVM-D1
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
Appendix 1. Electromagnetic Compatibility
Essential Performance
The product is designed to provide images for observation, excluding short-term automatically
recoverable reduction caused by electromagnetic interference.
The following cable information is given for EMC reference only.
Cable
Max. cable length
Shielded/unshielded
Qty.
Category
AC power cable
1.8 m
Unshielded
1 set
AC power supply
Adapter cable
1.2 m
Unshielded
1 set
DC power supply
HMDI-DVI cable
2.8 m
Shielded
1 set
Signaling
SDI cable
2.8 m
Shielded
1 set
Signaling
Endoscope connection cable A
2m
Shielded
1 set
Signaling
Endoscope connection cable B
2m
Shielded
1 set
Signaling
HDMI cable
2.8 m
Shielded
1 set
Signaling
D-SUB9 cable
2m
Unshielded
1 set
Signaling
REMOTE cable
1.5m
Unshielded
1 set
Signaling
Important information on electromagnetic compatibility (EMC)
The product shall be subjected to special precautions regarding EMC, and the EMC information
provided in its user's manual shall apply when using it. The product complies with IEC 60601-1-2:2020
in terms of immunity and emission. However, the following special considerations shall be observed:
The equipment with above-mentioned ESSENTIAL PERFORMANCE is intended used in Professional
healthcare facility environment, except for any places that are near the RF shielding room of the ME
system for magnetic resonance imaging (MRI) with high electromagnetic interference intensity.
Warning
It is required to avoid using the product next to or stacked with other equipment; Otherwise, improper
operation may be caused. If you have to do so, the product and other equipment should be observed
carefully to verify that they can operate properly.
The use of accessories, transducers or cables other than those specified or provided by the
manufacturer of the product may increase electromagnetic radiation or reduce electromagnetic
interference, and cause improper operation.
When using any portable radio frequency communication equipment (including antenna cables,
external antennas and other peripheral equipment), such equipment should be placed over 30 cm (12
inches) away from any part of the product, including the cables specified by the manufacturer;
Otherwise, the performance of such equipment may be impaired.