84
Operating
D
IAGNOSTIC
P
RO
™ Advantage and D
OSIMETRY
P
RO
™ Advantage
About digitizers and film sizes
This section describes i
mportant differences between the
D
IAGNOSTICPRO™ Advantage and DOSIMETRYPRO™ Advantage.
DIAGNOSTICPRO™ Advantage specifications
Nominal Pixels Spot size Line pairs Digitizing
DPI
resolution (14"x17" film) (µm) per mm speed
1k x 1.25k 1008 x 1124 340 75 1 12.2 sec
2k x 2.5k* 2002 x 2431 170 150 3 12.2 sec
4k x 5k 3990 x 4845 85 300 6 24.4 sec
Mammography film: 18cm x 24cm
4k x 5k 4104 x 5472 44 570 11 27.4 sec
*ACR Standard for Teleradiology Guidelines [Revision 35 (1998)] recommends 2.5 line pairs/mm minimum
Scan modes 32-bit mapped to 12-bit and 8-bit grayscale outputs
Film sizes Width:
8" to 14" (20.3cm to 35.6cm)
(manual feed) Length: 8" to 51"† (20.3cm to 129.5cm)
Thickness: 0.006" to 0.010" (0.15mm to 0.51mm)
†Longer films require user support during feeding, and a
scanning application that handles long films
Auto film feeder Standard 25-film capacity (mixed sizes, no presorting)
“Light box loading”: head-up, normal reading, left
justified
Width: 8" to 14" (20.3cm x 35.6cm)
Length 8" to 17" (20.3cm x 43.2cm)
Thickness: 0.006" to 0.010" (0.15mm to 0.51mm)