REMOTE OPERATION
PM1000+ Power Analyzer User Manual © Voltech 2010 Page 39
“:AVG?”
Returns status of averaging of results.
0 → Disabled (default)
1 → Enabled
":BLK:ENB”
Enables blanking
":BLK:DIS”
Disables blanking
":BLK?”
Returns current blanking setting.
0 Disabled
1 Enabled (default)
":CAL:DATE?”
Returns the date of last calibration.
":CFG:LOAD <value>”
Loads requested configuration <value>= 0
to 5, 0 = default.
":CFG:SAVE <value>”
Saves the specified configuration
<value>= 1 through 5.
":CFG:PRINT <value>"
Prints the specified configuration
<value>= 1 through 5.
":COM:RS2:BAUD <value>”
Sets the RS232 baud rate <value> =
9600, 19200 or 38400
":COM:RS2:BAUD?”
Returns the current baud rate setting
":COM:IEE:ADDR <value>”
Sets the IEEE488 (GPIB) address
<value> = 1 - 30
":COM:IEE:ADDR?”
Returns the current IEEE488 (GPIB)
address setting
“:COM:ETH:GATE?”
Returns Default Gateway currently in use.
“:COM:ETH:IP?
Returns the IP address currently in use.
“:COM:ETH:SUB?”
Returns Subnet Mask currently in use.
“:COM:ETH:MAC <value>”
Will set the MAC address. Accepts the
lower 24 bits as ASCII hex string. Ex.
<value> = 5AB2F1.