REMOTE OPERATION
Page 42 © Voltech 2010 PM1000+ Power Analyzer User Manual
Choose to include or exclude THD
Harmonic zero:
<value> = 0 → exclude
<value> = 1 → include
Select the THD reference:
<value> = 0 → fundamental
<value> = 1 → rms
Sets the low pass frequency filter state:
<value> = 0 -> Low Pass Filter Disabled
<value> = 1 -> Low Pass Filter Enabled
Returns the state of the low pass filter.
Selects either Manual Start Method or
Clock Start Method.
<value> = 0 → Manual Start Method
<value> = 1 → Clock Start Method
Starts integration when in Manual Start
Method. Requires integration mode
active, manual start selected and
integration not running.
Stops integration when in Manual Start
Method. Requires integration mode active,
manual start selected and integration
running.
Resets integration values. Requires
integration mode active and integration
not running.
":INT:CLK:TIME xx-xx-xxX"
Sets the start time for the integrator when
configured for Clock Start Method. Start
time sent in current PM1000+ time format.
xx-xx-xxX stands for hh-mm-ss
(uppercase „X‟ is not used) for 24hr time
format or hh-mm-ss(A or P) for AM/PM
time format.