REMOTE OPERATION
Page 44 © Voltech 2010 PM1000+ Power Analyzer User Manual
":MOD:BAL”
Sets ballast mode.
":MOD?”
Returns the current mode. 0 = Normal
1 = Ballast
2 = Inrush
3 = Standby
4 = Integrator
":MOD:INR:RNG <value>”
Set Inrush current range <value> = 1
through 6.
":MOD:INR:VRNG <value>”
Set Inrush voltage range <value> = 1
through 4.
":MOD:INR:CLR"
Clears the Apk value when in Inrush
mode.
":MOD:SBY:PER <value>"
Sets the user defined period of averaging
in Low Power Standby mode. <value>= 1
to 300 seconds.
":DSP:Z04”
Displays 4 results screen.
":DSP:Z14”
Displays 14 results screen.
":REM:OFF”
Returns PM1000+ from remote control.
":RNG:VLT:FIX <value>"
Fixes voltage range <value> = 1 (10V) to
4 (1000V)
":RNG:AMP:FIX <value>"
Fixes current range <value> = 1 (0.1A) to
6 (100A)
":RNG:VLT:AUT”
Sets voltage on auto range.
":RNG:AMP:AUT”
Sets current on auto range.
":RNG:VLT?”
Returns the current voltage range.
":RNG:AMP?”
Returns the current amps range.
":RNG:VLT:AUT?”
Returns : 0 → Range fixed.
1 → AutoRange engaged.
":RNG:AMP:AUT?”
Returns : 0 → Range fixed.
1 → AutoRange engaged.